{
    "dataType": "CVE_RECORD",
    "dataVersion": "5.2",
    "cveMetadata": {
        "cveId": "CVE-2025-14096",
        "assignerOrgId": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
        "state": "PUBLISHED",
        "assignerShortName": "Radiometer",
        "dateReserved": "2025-12-05T10:50:00.566Z",
        "datePublished": "2025-12-17T12:19:27.489Z",
        "dateUpdated": "2025-12-17T14:43:16.523Z"
    },
    "containers": {
        "cna": {
            "affected": [
                {
                    "defaultStatus": "unaffected",
                    "platforms": [
                        "Windows XP",
                        "Windows 7"
                    ],
                    "product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
                    "vendor": "Radiometer Medical Aps",
                    "versions": [
                        {
                            "status": "affected",
                            "version": "Windows 7 Operating system",
                            "versionType": "Radiometer Customized Windows Operating System"
                        },
                        {
                            "status": "affected",
                            "version": "Windows XP Operating system",
                            "versionType": "Radiometer Customized Windows Operating System"
                        }
                    ]
                },
                {
                    "defaultStatus": "unaffected",
                    "platforms": [
                        "Windows 7",
                        "Windows XP"
                    ],
                    "product": "AQT90 FLEX Analyzers",
                    "vendor": "Radiometer Medical Aps",
                    "versions": [
                        {
                            "status": "affected",
                            "version": "Windows 7 Operating system",
                            "versionType": "Radiometer Customized Windows Operating System"
                        },
                        {
                            "status": "affected",
                            "version": "Windows XP Operating system",
                            "versionType": "Radiometer Customized Windows Operating System"
                        }
                    ]
                },
                {
                    "defaultStatus": "unaffected",
                    "platforms": [
                        "Windows 7",
                        "Windows XP"
                    ],
                    "product": "ABL800 BASIC and ABL800 FLEX analyzers",
                    "vendor": "Radiometer Medical Aps",
                    "versions": [
                        {
                            "status": "affected",
                            "version": "Windows 7 Operating system",
                            "versionType": "Radiometer Customized Windows Operating System"
                        },
                        {
                            "status": "affected",
                            "version": "Windows XP Operating system",
                            "versionType": "Radiometer Customized Windows Operating System"
                        }
                    ]
                }
            ],
            "configurations": [
                {
                    "lang": "en",
                    "supportingMedia": [
                        {
                            "base64": false,
                            "type": "text/html",
                            "value": "Requires physical access to the Analyzer.\n\n<br>"
                        }
                    ],
                    "value": "Requires physical access to the Analyzer."
                }
            ],
            "credits": [
                {
                    "lang": "en",
                    "type": "reporter",
                    "value": "Florian Hauser and Fabian Weber from CODE WHITE GmbH"
                }
            ],
            "descriptions": [
                {
                    "lang": "en",
                    "supportingMedia": [
                        {
                            "base64": false,
                            "type": "text/html",
                            "value": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.<br><br>Other related CVE's are CVE-2025-14095 &amp; CVE-2025-14097.<br><br>\n\n<span style=\"background-color: rgb(255, 255, 255);\">Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.</span>\n\n<br><br>\n\n<span style=\"background-color: rgb(255, 255, 255);\">Required Configuration for Exposure:<br></span>\n\n<span style=\"background-color: rgb(255, 255, 255);\">Attacker requires physical access to the analyzer.</span>\n\n<br><br>Temporary work Around:<br>Only authorized people can physically access the analyzer.<br><br>Permanent solution:<br>Local Radiometer representatives will contact all affected customers to discuss a permanent solution.<br><br>Exploit Status:<br>\n\n<span style=\"background-color: rgb(255, 255, 255);\">Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.</span>\n\n<br><br>"
                        }
                    ],
                    "value": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\n\nOther related CVE's are CVE-2025-14095 & CVE-2025-14097.\n\n\n\nAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\n\n\nRequired Configuration for Exposure:\n\n\nAttacker requires physical access to the analyzer.\n\n\n\nTemporary work Around:\nOnly authorized people can physically access the analyzer.\n\nPermanent solution:\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\nExploit Status:\n\n\nResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."
                }
            ],
            "exploits": [
                {
                    "lang": "en",
                    "supportingMedia": [
                        {
                            "base64": false,
                            "type": "text/html",
                            "value": "Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\n\n<br>"
                        }
                    ],
                    "value": "Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."
                }
            ],
            "impacts": [
                {
                    "capecId": "CAPEC-69",
                    "descriptions": [
                        {
                            "lang": "en",
                            "value": "CAPEC-69 — Target Programs with Elevated Privileges"
                        }
                    ]
                }
            ],
            "metrics": [
                {
                    "cvssV3_1": {
                        "attackComplexity": "LOW",
                        "attackVector": "LOCAL",
                        "availabilityImpact": "HIGH",
                        "baseScore": 8.4,
                        "baseSeverity": "HIGH",
                        "confidentialityImpact": "HIGH",
                        "integrityImpact": "HIGH",
                        "privilegesRequired": "NONE",
                        "scope": "UNCHANGED",
                        "userInteraction": "NONE",
                        "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
                        "version": "3.1"
                    },
                    "format": "CVSS",
                    "scenarios": [
                        {
                            "lang": "en",
                            "value": "GENERAL"
                        }
                    ]
                }
            ],
            "problemTypes": [
                {
                    "descriptions": [
                        {
                            "cweId": "CWE-798",
                            "description": "CWE-798: Use of Hard-coded Credentials",
                            "lang": "en",
                            "type": "CWE"
                        }
                    ]
                },
                {
                    "descriptions": [
                        {
                            "cweId": "CWE-250",
                            "description": "CWE-250: Execution with Unnecessary Privileges",
                            "lang": "en",
                            "type": "CWE"
                        }
                    ]
                }
            ],
            "providerMetadata": {
                "orgId": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
                "shortName": "Radiometer",
                "dateUpdated": "2025-12-17T13:03:05.696Z"
            },
            "references": [
                {
                    "url": "https://www.radiometer.com/myradiometer"
                }
            ],
            "solutions": [
                {
                    "lang": "en",
                    "supportingMedia": [
                        {
                            "base64": false,
                            "type": "text/html",
                            "value": "Local Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\n\n\n<br>"
                        }
                    ],
                    "value": "Local Radiometer representatives will contact all affected customers to discuss a permanent solution."
                }
            ],
            "source": {
                "discovery": "EXTERNAL"
            },
            "timeline": [
                {
                    "lang": "en",
                    "time": "2025-12-17T13:01:00.000Z",
                    "value": "CVE Published"
                }
            ],
            "title": "Credential Disclosure vulnerability in Radiometer Products",
            "workarounds": [
                {
                    "lang": "en",
                    "supportingMedia": [
                        {
                            "base64": false,
                            "type": "text/html",
                            "value": "Only authorized people can physically access the analyzer.\n\n\n\n<br>"
                        }
                    ],
                    "value": "Only authorized people can physically access the analyzer."
                }
            ],
            "x_generator": {
                "engine": "Vulnogram 0.5.0"
            }
        },
        "adp": [
            {
                "metrics": [
                    {
                        "other": {
                            "type": "ssvc",
                            "content": {
                                "timestamp": "2025-12-17T14:42:57.843796Z",
                                "id": "CVE-2025-14096",
                                "options": [
                                    {
                                        "Exploitation": "none"
                                    },
                                    {
                                        "Automatable": "no"
                                    },
                                    {
                                        "Technical Impact": "total"
                                    }
                                ],
                                "role": "CISA Coordinator",
                                "version": "2.0.3"
                            }
                        }
                    }
                ],
                "title": "CISA ADP Vulnrichment",
                "providerMetadata": {
                    "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
                    "shortName": "CISA-ADP",
                    "dateUpdated": "2025-12-17T14:43:16.523Z"
                }
            }
        ]
    }
}