    "dataType": "CVE_RECORD",
    "dataVersion": "5.1",
    "cveMetadata": {
        "cveId": "CVE-2025-0683",
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "state": "PUBLISHED",
        "assignerShortName": "icscert",
        "dateReserved": "2025-01-23T18:11:20.770Z",
        "datePublished": "2025-01-30T18:17:35.867Z",
        "dateUpdated": "2025-02-12T20:41:36.789Z"
    },
    "containers": {
        "cna": {
            "affected": [
                {
                    "defaultStatus": "unaffected",
                    "product": "CMS8000 Patient Monitor",
                    "vendor": "Contec Health",
                    "versions": [
                        {
                            "status": "affected",
                            "version": "All versions"
                        }
                    ]
                }
            ],
            "credits": [
                {
                    "lang": "en",
                    "type": "finder",
                    "value": "An anonymous researcher reported these vulnerabilities to CISA."
                }
            ],
            "descriptions": [
                {
                    "lang": "en",
                    "supportingMedia": [
                        {
                            "base64": false,
                            "type": "text/html",
                            "value": "In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \npatient data to a hard-coded public IP address when a patient is hooked \nup to the monitor. This could lead to a leakage of confidential patient \ndata to any device with that IP address or an attacker in a \nmachine-in-the-middle scenario."
                        }
                    ],
                    "value": "In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \npatient data to a hard-coded public IP address when a patient is hooked \nup to the monitor. This could lead to a leakage of confidential patient \ndata to any device with that IP address or an attacker in a \nmachine-in-the-middle scenario."
                }
            ],
            "metrics": [
                {
                    "cvssV4_0": {
                        "Automatable": "NOT_DEFINED",
                        "Recovery": "NOT_DEFINED",
                        "Safety": "NOT_DEFINED",
                        "attackComplexity": "LOW",
                        "attackRequirements": "PRESENT",
                        "attackVector": "NETWORK",
                        "baseScore": 8.2,
                        "baseSeverity": "HIGH",
                        "privilegesRequired": "NONE",
                        "providerUrgency": "NOT_DEFINED",
                        "subAvailabilityImpact": "NONE",
                        "subConfidentialityImpact": "NONE",
                        "subIntegrityImpact": "NONE",
                        "userInteraction": "NONE",
                        "valueDensity": "NOT_DEFINED",
                        "vectorString": "CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N",
                        "version": "4.0",
                        "vulnAvailabilityImpact": "NONE",
                        "vulnConfidentialityImpact": "HIGH",
                        "vulnIntegrityImpact": "NONE",
                        "vulnerabilityResponseEffort": "NOT_DEFINED"
                    },
                    "format": "CVSS",
                    "scenarios": [
                        {
                            "lang": "en",
                            "value": "GENERAL"
                        }
                    ]
                },
                {
                    "cvssV3_1": {
                        "attackComplexity": "HIGH",
                        "attackVector": "NETWORK",
                        "availabilityImpact": "NONE",
                        "baseScore": 5.9,
                        "baseSeverity": "MEDIUM",
                        "confidentialityImpact": "HIGH",
                        "integrityImpact": "NONE",
                        "privilegesRequired": "NONE",
                        "scope": "UNCHANGED",
                        "userInteraction": "NONE",
                        "vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N",
                        "version": "3.1"
                    },
                    "format": "CVSS",
                    "scenarios": [
                        {
                            "lang": "en",
                            "value": "GENERAL"
                        }
                    ]
                }
            ],
            "problemTypes": [
                {
                    "descriptions": [
                        {
                            "cweId": "CWE-359",
                            "description": "CWE-359 Exposure of Private Personal Information to an Unauthorized Actor",
                            "lang": "en",
                            "type": "CWE"
                        }
                    ]
                }
            ],
            "providerMetadata": {
                "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
                "shortName": "icscert",
                "dateUpdated": "2025-01-31T17:07:47.786Z"
            },
            "references": [
                {
                    "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01"
                },
                {
                    "url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"
                }
            ],
            "solutions": [
                {
                    "lang": "en",
                    "supportingMedia": [
                        {
                            "base64": false,
                            "type": "text/html",
                            "value": "<p>Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.</p><p>Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's <a target=\"_blank\" rel=\"nofollow\" href=\"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication\">safety communication</a>.</p>\n\n<br>"
                        }
                    ],
                    "value": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\n\nPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's  safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication ."
                }
            ],
            "source": {
                "advisory": "ICSMA-25-030-01",
                "discovery": "EXTERNAL"
            },
            "title": "Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor",
            "x_generator": {
                "engine": "Vulnogram 0.2.0"
            }
        },
        "adp": [
            {
                "providerMetadata": {
                    "orgId": "af854a3a-2127-422b-91ae-364da2661108",
                    "shortName": "CVE",
                    "dateUpdated": "2025-01-31T15:35:10.389Z"
                },
                "references": [
                    {
                        "url": "https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor"
                    },
                    {
                        "url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"
                    },
                    {
                        "url": "https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/"
                    }
                ],
                "title": "CVE Program Container",
                "x_generator": {
                    "engine": "ADPogram 0.0.1"
                }
            },
            {
                "metrics": [
                    {
                        "other": {
                            "type": "ssvc",
                            "content": {
                                "id": "CVE-2025-0683",
                                "role": "CISA Coordinator",
                                "options": [
                                    {
                                        "Exploitation": "none"
                                    },
                                    {
                                        "Automatable": "no"
                                    },
                                    {
                                        "Technical Impact": "partial"
                                    }
                                ],
                                "version": "2.0.3",
                                "timestamp": "2025-01-30T19:06:59.525559Z"
                            }
                        }
                    }
                ],
                "title": "CISA ADP Vulnrichment",
                "providerMetadata": {
                    "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
                    "shortName": "CISA-ADP",
                    "dateUpdated": "2025-02-12T20:41:36.789Z"
                }
            }
        ]
    }
}