    "dataType": "CVE_RECORD",
    "dataVersion": "5.1",
    "cveMetadata": {
        "cveId": "CVE-2022-26388",
        "assignerOrgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
        "state": "PUBLISHED",
        "assignerShortName": "Baxter",
        "dateReserved": "2022-03-03T22:04:24.533Z",
        "datePublished": "2025-02-07T17:06:30.134Z",
        "dateUpdated": "2025-02-07T18:50:58.812Z"
    },
    "containers": {
        "cna": {
            "affected": [
                {
                    "defaultStatus": "unaffected",
                    "product": "ELI 380 Resting Electrocardiograph",
                    "vendor": "Welch Allyn",
                    "versions": [
                        {
                            "lessThanOrEqual": "2.6.0",
                            "status": "affected",
                            "version": "0",
                            "versionType": "custom"
                        }
                    ]
                },
                {
                    "defaultStatus": "unaffected",
                    "product": "ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph",
                    "vendor": "Welch Allyn",
                    "versions": [
                        {
                            "lessThanOrEqual": "2.3.1",
                            "status": "affected",
                            "version": "0",
                            "versionType": "custom"
                        }
                    ]
                },
                {
                    "defaultStatus": "unaffected",
                    "product": "ELI 250c/BUR 250c Resting Electrocardiograph",
                    "vendor": "Welch Allyn",
                    "versions": [
                        {
                            "lessThanOrEqual": "2.1.2",
                            "status": "affected",
                            "version": "0",
                            "versionType": "custom"
                        }
                    ]
                },
                {
                    "defaultStatus": "unaffected",
                    "product": "ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph",
                    "vendor": "Welch Allyn",
                    "versions": [
                        {
                            "lessThanOrEqual": "2.2.0",
                            "status": "affected",
                            "version": "0",
                            "versionType": "custom"
                        }
                    ]
                }
            ],
            "credits": [
                {
                    "lang": "en",
                    "type": "finder",
                    "value": "An anonymous user reported these vulnerabilities to Hillrom."
                }
            ],
            "descriptions": [
                {
                    "lang": "en",
                    "supportingMedia": [
                        {
                            "base64": false,
                            "type": "text/html",
                            "value": "A use of hard-coded password vulnerability may allow authentication abuse.<p>This issue affects ELI 380 Resting Electrocardiograph: \n\n<span style=\"background-color: rgb(255, 255, 255);\">Versions 2.6.0 and prior</span>; ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph:\n\n<span style=\"background-color: rgb(255, 255, 255);\">Versions 2.3.1 and prior</span>; ELI 250c/BUR 250c Resting Electrocardiograph:\n\n<span style=\"background-color: rgb(255, 255, 255);\">Versions 2.1.2 and prior</span>; ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: \n\n<span style=\"background-color: rgb(255, 255, 255);\">Versions 2.2.0 and prior</span>.</p>"
                        }
                    ],
                    "value": "A use of hard-coded password vulnerability may allow authentication abuse.This issue affects ELI 380 Resting Electrocardiograph: \n\nVersions 2.6.0 and prior; ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph:\n\nVersions 2.3.1 and prior; ELI 250c/BUR 250c Resting Electrocardiograph:\n\nVersions 2.1.2 and prior; ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: \n\nVersions 2.2.0 and prior."
                }
            ],
            "impacts": [
                {
                    "capecId": "CAPEC-114",
                    "descriptions": [
                        {
                            "lang": "en",
                            "value": "CAPEC-114 Authentication Abuse"
                        }
                    ]
                }
            ],
            "metrics": [
                {
                    "cvssV3_1": {
                        "attackComplexity": "LOW",
                        "attackVector": "PHYSICAL",
                        "availabilityImpact": "LOW",
                        "baseScore": 6.4,
                        "baseSeverity": "MEDIUM",
                        "confidentialityImpact": "HIGH",
                        "integrityImpact": "HIGH",
                        "privilegesRequired": "NONE",
                        "scope": "UNCHANGED",
                        "userInteraction": "NONE",
                        "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:L",
                        "version": "3.1"
                    },
                    "format": "CVSS",
                    "scenarios": [
                        {
                            "lang": "en",
                            "value": "GENERAL"
                        }
                    ]
                }
            ],
            "problemTypes": [
                {
                    "descriptions": [
                        {
                            "cweId": "CWE-259",
                            "description": "CWE-259 Use of Hard-Coded Password",
                            "lang": "en",
                            "type": "CWE"
                        }
                    ]
                }
            ],
            "providerMetadata": {
                "orgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
                "shortName": "Baxter",
                "dateUpdated": "2025-02-07T17:06:30.134Z"
            },
            "references": [
                {
                    "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-22-167-01"
                },
                {
                    "url": "https://hillrom.com/en/responsible-disclosures/"
                }
            ],
            "solutions": [
                {
                    "lang": "en",
                    "supportingMedia": [
                        {
                            "base64": false,
                            "type": "text/html",
                            "value": "<p>Hillrom has released software updates for all impacted devices to \naddress these vulnerabilities. New product versions that mitigate these \nvulnerabilities are available as follows:</p>\n<ul>\n<li>Welch Allyn ELI 380 Resting Electrocardiograph: available Q4 2023</li>\n<li>Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: available May 2022</li>\n<li>Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: available Q4 2023</li>\n</ul>\n\n\nHillrom recommends users upgrade to the latest product versions. \nInformation on how to update these products can be found on the <a target=\"_blank\" rel=\"nofollow\" href=\"https://hillrom.com/en/responsible-disclosures/\">Hillrom disclosure page</a>.\n\n<br>"
                        }
                    ],
                    "value": "Hillrom has released software updates for all impacted devices to \naddress these vulnerabilities. New product versions that mitigate these \nvulnerabilities are available as follows:\n\n\n\n  *  Welch Allyn ELI 380 Resting Electrocardiograph: available Q4 2023\n\n  *  Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: available May 2022\n\n  *  Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: available Q4 2023\n\n\n\n\n\n\nHillrom recommends users upgrade to the latest product versions. \nInformation on how to update these products can be found on the  Hillrom disclosure page https://hillrom.com/en/responsible-disclosures/ ."
                }
            ],
            "source": {
                "advisory": "ICSMA-22-167-01",
                "discovery": "EXTERNAL"
            },
            "title": "Use of Hard-Coded Password Vulnerability in ELI Electrocardiograph Devices",
            "workarounds": [
                {
                    "lang": "en",
                    "supportingMedia": [
                        {
                            "base64": false,
                            "type": "text/html",
                            "value": "<p>Hillrom recommends the following workarounds to help reduce risk:</p>\n<ul>\n<li>Apply proper network and physical security controls.</li>\n<li>Ensure a unique encryption key is configured for ELI Link and Cardiograph.</li>\n<li>Where possible, use a firewall to prevent communication on Port 21 \nFTP service, Port 22 SSH (Secure Shell Connection), and Port 23 Telnet \nservice.</li>\n</ul>\n\n<br>"
                        }
                    ],
                    "value": "Hillrom recommends the following workarounds to help reduce risk:\n\n\n\n  *  Apply proper network and physical security controls.\n\n  *  Ensure a unique encryption key is configured for ELI Link and Cardiograph.\n\n  *  Where possible, use a firewall to prevent communication on Port 21 \nFTP service, Port 22 SSH (Secure Shell Connection), and Port 23 Telnet \nservice."
                }
            ],
            "x_generator": {
                "engine": "Vulnogram 0.2.0"
            }
        },
        "adp": [
            {
                "metrics": [
                    {
                        "other": {
                            "type": "ssvc",
                            "content": {
                                "timestamp": "2025-02-07T18:50:50.198188Z",
                                "id": "CVE-2022-26388",
                                "options": [
                                    {
                                        "Exploitation": "none"
                                    },
                                    {
                                        "Automatable": "no"
                                    },
                                    {
                                        "Technical Impact": "total"
                                    }
                                ],
                                "role": "CISA Coordinator",
                                "version": "2.0.3"
                            }
                        }
                    }
                ],
                "title": "CISA ADP Vulnrichment",
                "providerMetadata": {
                    "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
                    "shortName": "CISA-ADP",
                    "dateUpdated": "2025-02-07T18:50:58.812Z"
                }
            }
        ]
    }
}