synced 12 MIN AGO
01 Record

CVE-2021-44464

Fresenius Kabi Agilia Connect Infusion System hard coded credentials

MediumPUBLISHEDCNA: icscert
CNA base score6.3 CVSS v3.xCVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:L/A:L

02 Description

Fresenius Kabi Agilia Connect Infusion System hard coded credentials

03 Exploitation and scoring

Exploitation
Assessed, none found. CISA's assessment recorded neither public exploit code nor evidence of exploitation when it was made.
SSVC decision
  • ExploitationnoneCISA recorded no public exploit code and no evidence of exploitation.
  • AutomatablenoAt least one step from reconnaissance to exploitation cannot be reliably automated.
  • Technical ImpactpartialSuccessful exploitation gives limited control of the vulnerable component.
CISA publishes the decision points, not a final SSVC decision. The decision also depends on mission and well-being impact, which is a property of your deployment rather than of the vulnerability.
EPSS probability
0.64% probability of exploitation activity in the next 30 days.
EPSS percentile
Equal to or higher than the score of 50.5% of scored records. A rank, not a probability.
NVD base score
8.8 (CVSS v3.x), scored by the NVD independently of the CNA.
NVD CVSS vector
CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H
NVD analysis status
Modified. NVD last modified this record on 2026-06-17.
Red Hat severity
No Red Hat rating. Red Hat rates the CVEs that affect its products.

Exploit-prediction scores from FIRST. Independent scoring and analysis status from the NVD. Vendor severity from Red Hat. Exploitation assessment and catalog membership from CISA.

04 Metadata

Published
2022-01-21 18:17Z4 YR AGO
Last updated
2025-04-16 16:47Z1 YR AGO
Reserved
2021-11-30
Assigning CNA
icscert
Record state
PUBLISHED
Severity
Medium (CVSS 6.3)
CNA CVSS vector
CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:L/A:L
Weaknesses
  • CWE-798Use of Hard-coded CredentialsBase
Data version
5.1
Document digest
8a39424eac2b3ad3c0439613c815b4bf0e229ab4cb8c217555beaa08985e06ac

05 Affected products

VendorProductVersionsPlatforms

06 References

07 Remediation

Fresenius Kabi has created new versions to address these vulnerabilities: • Link+ v3.0 (D16 or later) • VSS v1.0.3 (or later) • Agilia Connect Pumps Wifi Module (D29 or later) • Agilia Connect Partner v3.3.2 (or later) Fresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version. Fresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers’ installation, Fresenius Kabi recommends users to rely on CISA recommendation.