CVE-2021-43355
Fresenius Kabi Agilia Connect Infusion System use of client side authentication
02 Description
Fresenius Kabi Agilia Connect Infusion System use of client side authentication
03 Exploitation and scoring
- Exploitation
- Assessed, none found. CISA's assessment recorded neither public exploit code nor evidence of exploitation when it was made.
- SSVC decision
- ExploitationnoneCISA recorded no public exploit code and no evidence of exploitation.
- AutomatableyesAn attacker can reliably automate reconnaissance through exploitation, so this scales.
- Technical ImpactpartialSuccessful exploitation gives limited control of the vulnerable component.
- EPSS probability
- 0.97% probability of exploitation activity in the next 30 days.
- EPSS percentile
- Equal to or higher than the score of 61.7% of scored records. A rank, not a probability.
- NVD base score
- 9.8 (CVSS v3.x), scored by the NVD independently of the CNA.
- NVD CVSS vector
- CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- NVD analysis status
- Modified. NVD last modified this record on 2026-06-17.
- Red Hat severity
- No Red Hat rating. Red Hat rates the CVEs that affect its products.
Exploit-prediction scores from FIRST. Independent scoring and analysis status from the NVD. Vendor severity from Red Hat. Exploitation assessment and catalog membership from CISA.
04 Metadata
- Published
- 2022-01-21 18:17Z4 YR AGO
- Last updated
- 2025-04-16 16:46Z1 YR AGO
- Reserved
- 2021-11-30
- Assigning CNA
- icscert
- Record state
- PUBLISHED
- Severity
- High (CVSS 7.3)
- CNA CVSS vector
- CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L
- Weaknesses
- Data version
- 5.1
- Document digest
- fdfd263fc2ddf34ca6187b784d5a616f8694e665fefeea11393375af83494d19
05 Affected products
06 References
07 Remediation
Fresenius Kabi has created new versions to address these vulnerabilities: • Link+ v3.0 (D16 or later) • VSS v1.0.3 (or later) • Agilia Connect Pumps Wifi Module (D29 or later) • Agilia Connect Partner v3.3.2 (or later) Fresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version. Fresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers’ installation, Fresenius Kabi recommends users to rely on CISA recommendation.