synced 3 MIN AGO
01 Record

CVE-2019-6538

Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control

CriticalPUBLISHEDCNA: icscert
CNA base score9.3 CVSS v3.xCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H

02 Description

Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control

03 Exploitation and scoring

Exploitation
Not assessed. CISA has not published an exploitation assessment for this record, and it is not in the KEV catalog. That is an absence of assessment, not a finding of no risk.
SSVC decision
Not assessed by CISA
EPSS probability
0.89% probability of exploitation activity in the next 30 days.
EPSS percentile
Equal to or higher than the score of 59.1% of scored records. A rank, not a probability.
NVD base score
6.5 (CVSS v3.x), scored by the NVD independently of the CNA.
NVD CVSS vector
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
NVD analysis status
Modified. NVD last modified this record on 2026-06-17.
Red Hat severity
No Red Hat rating. Red Hat rates the CVEs that affect its products.

Exploit-prediction scores from FIRST. Independent scoring and analysis status from the NVD. Vendor severity from Red Hat. Exploitation assessment and catalog membership from CISA.

04 Metadata

Published
2019-03-25 21:26Z7 YR AGO
Last updated
2025-05-22 19:26Z1 YR AGO
Reserved
2019-01-22
Assigning CNA
icscert
Record state
PUBLISHED
Severity
Critical (CVSS 9.3)
CNA CVSS vector
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H
Weaknesses
  • CWE-284Improper Access ControlPillar
  • CWE-306Missing Authentication for Critical FunctionBase
  • CWE-862Missing AuthorizationClass
Data version
5.1
Document digest
ff31e780db6e7bf1bdb5e8b064a1739b3b85388692cab3ba96ee87407701bd19

05 Affected products

VendorProductVersionsPlatforms
MedtronicMyCareLink Monitor24950, 24952—
MedtronicCareLink Monitor2490C—
MedtronicCareLink 2090 ProgrammerAll versions—
MedtronicAmplia CRT-DAll versions—
MedtronicClaria CRT-DAll versions—
MedtronicCompia CRT-DAll versions—
MedtronicConcerto CRT-DAll versions—
MedtronicConcerto II CRT-DAll versions—
MedtronicConsulta CRT-DAll versions—
MedtronicEvera ICDAll versions—
MedtronicMaximo II CRT-DAll versions—
MedtronicMaximo II ICDAll versions—
MedtronicMirro ICDAll versions—
MedtronicNayamed ND ICDAll versions—
MedtronicPrimo ICDAll versions—
MedtronicProtecta ICD, Protecta CRT-DAll versions—
MedtronicSecura ICDAll versions—
MedtronicVirtuoso ICDAll versions—
MedtronicVirtuoso II ICDAll versions—
MedtronicVisia AF ICDAll versions—
MedtronicViva CRT-DAll versions—
MedtronicBrava CRT-DAll versions—
MedtronicMirro MRI ICDAll versions—

06 References

07 Remediation

Medtronic has developed mitigating patches for a subset of the affected implanted cardiac device models. These patches are installed during regular office visits. Medtronic has stated that patches for additional impacted models are being developed by Medtronic and will be deployed through future updates. Patches are currently available for the following affected models: * Protecta CRT-D and implantable cardioverter defibrillators (ICDs), all models * Amplia MRI CRT-D, all models (patch available in U.S. only) * Claria MRI CRT-D, all models (patch available in U.S. only) * Compia MRI CRT-D, all models (patch available in U.S. only) * Visia AF MRI ICD, all models * Visia AF ICD, all models * Brava CRT-D, all models * Evera MRI ICD, all models * Evera ICD, all models * Mirro MRI ICD, all models * Primo MRI ICD, all models * Viva CRT-D, all models