synced JUST NOW
01 Record

CVE-2019-13543

Medtronic Valleylab FT10 and FX8 Use of Hard-coded Credentials

MediumPUBLISHEDCNA: icscert
CNA base score5.8 CVSS v3.xCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:L/I:N/A:N

02 Description

Medtronic Valleylab FT10 and FX8 Use of Hard-coded Credentials

03 Exploitation and scoring

Exploitation
Not assessed. CISA has not published an exploitation assessment for this record, and it is not in the KEV catalog. That is an absence of assessment, not a finding of no risk.
SSVC decision
Not assessed by CISA
EPSS probability
2.01% probability of exploitation activity in the next 30 days.
EPSS percentile
Equal to or higher than the score of 80.7% of scored records. A rank, not a probability.
NVD base score
7.5 (CVSS v3.x), scored by the NVD independently of the CNA.
NVD CVSS vector
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
NVD analysis status
Modified. NVD last modified this record on 2026-06-17.
Red Hat severity
No Red Hat rating. Red Hat rates the CVEs that affect its products.

Exploit-prediction scores from FIRST. Independent scoring and analysis status from the NVD. Vendor severity from Red Hat. Exploitation assessment and catalog membership from CISA.

04 Metadata

Published
2019-11-08 19:03Z6 YR AGO
Last updated
2025-05-22 19:03Z1 YR AGO
Reserved
2019-07-11
Assigning CNA
icscert
Record state
PUBLISHED
Severity
Medium (CVSS 5.8)
CNA CVSS vector
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:L/I:N/A:N
Weaknesses
  • CWE-798Use of Hard-coded CredentialsBase
Data version
5.1
Document digest
b1c0c380544725178bb83d79fbaf39a92c7c63287bd9d09d60b1ba4620836aac

05 Affected products

VendorProductVersionsPlatforms
MedtronicValleylab Exchange Client0 to 3.4—
MedtronicValleylab FT10 Energy Platform (VLFT10GEN)0 to software version 4.0.0—
MedtronicValleylab FX8 Energy Platform (VLFX8GEN)0 to software version 1.1.0—

06 References

07 Remediation

Software patches are currently available for the FT10 platform and will be available in early 2020 for the FX8 platform. Until these updates can be applied, Medtronic recommends to either disconnect affected products from IP networks or to segregate those networks, such that the devices are not accessible from an untrusted network (e.g., Internet). Patches can be downloaded at the following location: https://www.medtronic.com/covidien/en-us/support/software.html Medtronic has released additional patient focused information, at the following location: https://www.medtronic.com/security