CVE-2014-5406
Hospira LifeCare PCA Infusion System
02 Description
Hospira LifeCare PCA Infusion System
03 Exploitation and scoring
- Exploitation
- Not assessed. CISA has not published an exploitation assessment for this record, and it is not in the KEV catalog. That is an absence of assessment, not a finding of no risk.
- SSVC decision
- Not assessed by CISA
- EPSS probability
- 1.21% probability of exploitation activity in the next 30 days.
- EPSS percentile
- Equal to or higher than the score of 68.1% of scored records. A rank, not a probability.
- NVD base score
- The NVD has published no score for this record
- NVD CVSS vector
- No NVD score to derive
- NVD analysis status
- Modified. NVD last modified this record on 2026-06-17.
- Red Hat severity
- No Red Hat rating. Red Hat rates the CVEs that affect its products.
Exploit-prediction scores from FIRST. Independent scoring and analysis status from the NVD. Vendor severity from Red Hat. Exploitation assessment and catalog membership from CISA.
04 Metadata
- Published
- 2015-07-06 19:10Z11 YR AGO
- Last updated
- 2025-11-03 18:34Z11 MO AGO
- Reserved
- 2014-08-22
- Assigning CNA
- icscert
- Record state
- PUBLISHED
- Severity
- Not rated by the issuing CNA
- CNA CVSS vector
- Not published by the issuing CNA
- Weaknesses
- CWE-345Insufficient Verification of Data AuthenticityClass
- Data version
- 5.2
- Document digest
- 693ebc9de0c181e5dac4511c7a257c057448cf2b0982331b6d81437b25bfcada
05 Affected products
06 References
- www.cisa.govhttps://www.cisa.gov/news-events/ics-advisories/icsa-15-125-01
- www.fda.govhttp://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm
- xs-sniper.comhttps://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/
- github.comhttps://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2015/icsa-15-125-01.json
- ics-cert.us-cert.govhttps://ics-cert.us-cert.gov/advisories/ICSA-15-125-01
07 Remediation
ICS-CERT has been working with Hospira since May 2014 to address the vulnerabilities in the LifeCare PCA Infusion System. Hospira has developed a new version of the PCS Infusion System, Version 7.0 that addresses the identified vulnerabilities. According to Hospira, Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access. Existing PCA Infusion Systems running Version 5.0 can be upgraded to Version 7.0 when it becomes available. Hospira’s Version 7.0 is being reviewed by the FDA prior to its release. The release date for Version 7.0 of the LifeCare PCA Infusion System has not been determined. For additional information about Hospira’s new release, contact Hospira’s technical support at 1‑800-241-4002.